Getting Started Manual
WR6Zi-GSM-E
130
4.
Touch the
Summary View
checkbox to enable a line of text (as
follows) showing a concise status of the last waveform and
keeping a running count of how many sweeps have passed.
5.
Touch inside the
Pass If
control, and select a Boolean condition
from the choices shown.
6.
If you want to set up the test to end after a finite number of
sweeps, touch the
Stop Test
checkbox. Then touch inside the
After
data entry control and provide a value using your preferred
input control method - for more information regarding the touch
screen controls, refer to
Dialog Area
(on page 46).
7.
Use either the
Pass
or
Fail
button on the
If
section of the dialog to
set actions taking place when your waveform passes or fails the
test.
8.
The
Then
section of the dialog provides
stop test
,
sound alarm
,
print result
,
emit pulse
, or
save
(the waveform) actions. If you
want to have the results printed (and your oscilloscope is not
equipped with a printer) make sure your instrument is connected
to a local or network printer. Learn more in the
Printing and
Hardcopy FunctionsPrinting and Hardcopy Functions
topic
(available in the online help files).
9.
If you want to save your waveform automatically, touch the
Save
Setup...
checkbox. This takes you out of the current dialog and
opens the
Save Waveform
dialog. Learn more in the
Save Recall
→ Saving and Recalling WaveformsSaving and Recalling
Waveforms
topic (available in the online help files).
10.
Test Pass/Fail conditions by touching the
Force Actions Once
button. Press the
Clear All
button to quickly uncheck all
checkboxes and make new selections.