![LeCroy WaveRunner 6Zi Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372032.webp)
Getting Started Manual
WR6Zi-GSM-E
22
READY and TRIG'D Indicators
- The READY indicator is lit when the
trigger is armed. TRIG'D is lit momentarily when a trigger occurs. A
fast trigger rate causes the light to stay lit continuously.
Setup
- Press once to open the
Trigger Setup...
dialog.
Corresponds with screen menu selection:
Trigger → Trigger
Setup...
. Press the Trigger S
ETUP
front panel button again to close
the
Trigger Setup...
dialog.
Trigger Scan
- Pressing the Trigger Scan button opens the
TriggerScan dialog to help you quickly find rare waveform glitches
and anomalies.
Source
- This button directly selects through the available
Source
field values (C1, C2, C3, C4, Dig, EXT, and Line) usually accessed on
the
Setup
section of the main
Trigger
dialog.
Slope
- This button directly selects through the available
Slope
field values (Positive, Negative, or Either) usually accessed on the
Setup
section of the main
Trigger
dialog.
Coupling
- This button directly selects through the available
Coupling
field values - DC, AC, LFR (Low-Frequency Rejection), or
HFR (High-Frequency Rejection) - usually accessed on the
Setup
section of the main
Trigger
dialog.
Auto
- Press to turn on
Auto Trigger
mode, which triggers the
oscilloscope after a time-out, even if the trigger conditions are not
met.
Normal
- Press to turn on
Normal Trigger
mode, which triggers
the oscilloscope each time a signal is present that meets the
conditions set for the type of trigger selected.
Single
- Press to turn on
Single Trigger
mode, which arms the
oscilloscope to trigger once (single-shot acquisition) when the
input signal meets the trigger conditions set for the type of trigger
selected. If the scope is already armed, it will force a trigger.
Stop
- Press to prevent the scope from triggering on a signal. If
you boot up the instrument with the trigger in
Stop
mode, a
no
trace available
message is shown. Press the Trigger A
UTO
front
panel button to display your trace.