![LeCroy WaveRunner 6Zi Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/lecroy/waverunner-6zi/waverunner-6zi_getting-started-manual_1866372085.webp)
WaveRunner 6 Zi Oscilloscopes
75
WR6Zi-GSM-E
MulitStage
Touching this
MultiStage
trigger type shows the additional MultiStage
types on the dialog. Select from
Cascaded
,
QualFirst
, and
Qualified
as
follows:
C
ASCADED
The Cascaded trigger allows you to define successive trigger
Stages
(referred to as Stage
A
,
B
,
C
, and
D
) as
arm
,
trigger/arm
,
trigger/arm
,
trigger
criteria, respectively.
Q
UAL
F
IRST
In single trigger mode,
QualFirst
arms the oscilloscope on the A event,
and then triggers on all subsequent B events.
Note
: This button is enabled when using the sequence sampling mode.
It is commonly used in sequence mode for disk drive applications with
the index pulse defined as the A qualifier signal and the servo gate
signal as the B triggering events.
Q
UALIFIED
A-B
In single trigger mode,
Qualify A-B
arms the oscilloscope on the A event,
and then triggers on the B event. If the oscilloscope is in Normal trigger
mode, it automatically resets after the B event. Arm trigger events can
be set for
Edge
,
Pattern
,
State
, and
PatState
. State is briefly explained as
follows: