Getting Started Manual
WR6Zi-GSM-E
176
As you change the gain, this control allows you to either keep the
vertical offset level indicator stationary (when
Div
is selected) or
to have it move with the actual voltage level (when
Volts
is
selected). The advantage of selecting
Div
is that the waveform
remains on the grid as you increase the gain; whereas, if
Volts
is
selected, the waveform could move off the grid.
Note
: Regardless of whether you select
Volts
or
Div
, the
Offset
shown in the channel setup dialog always indicates volts.
However, when
Div
is selected for the Offset Control, the offset
in volts is scaled proportional to the change in gain, thereby
keeping the division on the grid constant.
2.
On the
Delay Setting constant in
area of the dialog, touch either
the
Time
or
Div
button.
As you change the timebase, this control allows you to either keep
the horizontal offset indicator stationary (when
Div
is selected) or
to have it move with the trigger point (when
Time
is selected).
The advantage of selecting
Div
is that the trigger point remains on
the grid as you increase the timebase; whereas, if
Time
is
selected, the trigger point could move off the grid.
Note
: Regardless of whether you select
Time
or
Div
, the
Delay
shown in the timebase setup dialog always indicates time.
However, when
Div
is selected for Delay In, the delay in time is
scaled proportional to the change in timebase, thereby keeping
the division on the grid constant.
3.
If this checkbox control is available on your oscilloscope, you can
mark the
Reset trigger counter before starting a new acquisition
checkbox on the
Trigger Counter Setting
area of the dialog.