LeCroy Corporation
External Trig Setting
262
Sierra M6
‐
2 SAS/SATA Protocol Analyzer User Manual
External Trig Setting
The External Trig Setting dialog displays the External Trig Out Setting and External Trig In
Setting as High Active, Low Active, or Toggle.
.
Figure 3.83: External Trigger Setting Dialog
To display the External Trig Setting dialog, select
Configuration
>
External
Trig
Setting
.
External Trig Out Setting
The Analyzer can send a Low or High external signal anytime a trigger occurs. Select the
External Trig Out Setting: High Active, Low Active, or Toggle from High to Low or Low to
High once (3.3 V output).
Enter the External TrigOut pulse width.
External Trig In Setting
An external Low or High input signal can cause triggering. Select the
External Trig In Setting: High Active, Low Active, or Toggle from High to Low or Low to
High once (3.3 V output).
Note:
The nominal External Trigger voltage is 0.818 volts. Trigger In can work with 1 volt to 5 volts
input voltage.
Содержание Sierra M6-2
Страница 18: ...LeCroy Corporation Contents 16 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 42: ...LeCroy Corporation Viewing Captured Data 40 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 192: ...LeCroy Corporation Target and Device Emulation 190 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 241: ...Sierra M6 2 SAS SATA Protocol Analyzer User Manual 239 Search LeCroy Corporation Figure 3 64 SAS Search Data Pattern ...
Страница 314: ...LeCroy Corporation Running Verification Script Engine VSE 312 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 462: ...LeCroy Corporation Resetting the Toolbar 460 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Страница 468: ...LeCroy Corporation 466 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...