12
–7
Mass Storage Utilities
When “Mass Storage Utilities” is selected from “UTILITIES”
the “MASS STORAGE” menu group appears (
12–11
) to give
access to the mass-storage file system controls. The
system supports storage and retrieval of data files to and
from memory cards, floppy disks and removable hard disk
(HDD) media.
Memory Card Format
The Memory Card structure, based on the PCMCIA II / JEIDA 4.0
standard, and as found in any DOS floppy or hard disk, consists of a
DOS partition containing files. The oscilloscope formats the card in
segmented contiguous sectors, each of 512 bytes. The scope does
not
support error-detection algorithms such as CRCs or checksum
inserted between the sectors, and when this is done the instrument
may only be able to read, but not write, the card.
Floppy Disk Format
The floppy supports DOS 1.44 MB and 720 kB formats.
Hard Disk (HDD) Format
The remov able hard disk structure is based on the PCMCIA III /
JEIDA 4.0 standard. The media is arranged as a DOS partition
containing files as in any DOS floppy or hard disk.
The HDD format uses 512 bytes per sector and four sectors per
cluster. One cluster is the minimum file size: any files smaller than
2048 bytes in size will still use one cluster’s allocation of 2048 bytes
of disk space.
Subdirectories
All files are written to and read from the media from the current
working directory. The default name of the working directory is
LECROY_1.DIR. This directory is automatically created when the
media is formatted. If the media is formatted elsewhere — for
instance on a PC — the directory will be created the first time a file
is stored to the memory card, floppy disk or remov able hard disk.
The working directory can be changed to any v alid DOS directory
name, using the file-name preferences menu. All working directories
are created as sub-directories from the root directory.
The maximum number of files allowed in any one directory is 2400.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...