12
–23
CAL BNC Setup
CAL BNC OUT
When “CAL BNC Setup” is selected from “UTILITIES”,
selection can be made of the type of signal put out at the CAL
BNC connector. The frequency, amplitude and pulse shape of
the calibration signal can also be chosen.
In addition, the CAL BNC connector can be used to prov ide a pulse:
Ø
as an action for PASS/FAIL testing
Ø
at the occurrence of each accepted trigger ev ent (Trigger Out)
Ø
when the scope is ready to accept a trigger ev ent (Trigger Rdy).
When the instrument is switched on, the calibration signal is
automatically set to its default state, 1 kHz 1 V square wav e.
mode
To change the kind of signal.
SET TO
To quickly reset the CAL BNC output to its default state.
Shape
To change the form of the calibration signal.
Amplitude
Using the associated knob, for setting the desired high lev el for all
CAL BNC applications. If the BNC output is connected to an input
channel with 50
Ω
, the amplitude will be halv ed.
Frequency
Using the associated knob, for setting the desired frequency of a
CAL signal in the range 500 Hz–2 MHz.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...