–20
Triggers and When to Use Them
More About Pattern Trigger
Once the pattern is defined, one of two transitions can be used to
generate the trigger. When the pattern begins, called
entering
the
pattern, a trigger can be generated. Alternatively, a trigger can be
generated when the pattern ends,
exiting
the pattern.
With pattern triggering, as in single source, either of these
qualifications can be selected: Hold-off for 10 ns up to 20 s, or
Hold-off for 1 to 99 999 999 events.
Set to Pattern Trigger, the oscilloscope always checks the logic
AND of the defined input logic states. However, with the help of
de Morgan's laws, the pattern becomes far more generalized.
Consider the important example of the Bi-level or Window Pattern
Trigger. Bi-level implies the expectation of a single-shot signal on
which the amplitude will go in either direction outside a known
range. To set up a Bi-level Pattern trigger, connect the signal to two
inputs: Channels 1 and 2 or any other pair that can be triggered on.
For example, the threshold of CH 1 could be set to +100 mV and
that of CH 2 at
−
200 mV. The Bi-level Trigger will occur if the
oscilloscope triggers on CH 1 for any pulse greater than +100 mV,
or on CH 2 for any pulse less than –200 mV. For improved
precision, the gains of the two channels should be at the same
setting.
In Boolean notation we can write:
Trigger
CH 1 CH 2
=
+
,
i.e. trigger when entering the pattern CH 1 = high OR CH 2 = low.
By de Morgan's laws this is equivalent to:
Trigger
CH 1 CH 2
=
⋅
,
i.e. trigger when exiting the pattern CH 1 = low AND CH 2 = high.
This configuration can be easily programmed.
The possibility of setting the threshold individually for each channel
extends this method so that it becomes a more general Window
Trigger: in order to trigger the input pulse amplitude must lie within
or outside a given arbitrary window.
Pattern Trigger has been designed to allow a choice of the trigger
point. By choosing
1L*2H entering
, the trigger will occur at the
moment the pattern 1L*2H
becomes true.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...