6
–3
SNGL
Pressing this button places the scope in Single-Shot Mode,
where it waits for a single trigger to occur, then displays the
signal and stops acquiring. If no signal occurs, the button can be
pressed again to show the signal being observed without a
trigger.
Press SNGL when in RIS Mode and the instrument will wait for
all the trigger events required to build up one signal on screen
before it stops. This may require as many as 4000 trigger
events.
Single-Shot Roll Mode behavior is the same as standard Single-
Shot but without the need to press the button a second time to
show the signal.
DELAY
— is used to adjust the pre- or post-trigger delay. Pre-trigger
adjustment is available from zero to 100 % of the full time-scale
in steps of 1 %. The pre-trigger delay is illustrated by the vertical
arrow symbol at the bottom of the grid. Post-trigger adjustment
is available from 0 to 10 000 divisions in increments of 0.1 of a
division. The post-trigger-delay value is labeled in seconds and
is located in the on-screen Trigger Delay field.
ZERO
— sets the trigger delay at zero, the trigger instant at the left-
hand edge of the grid.
TIME/DIV
— selects the time per division in a 1–2–5 sequence. The
time/div setting is displayed in the Acquisition Summary field.
LEVEL
— adjusts the trigger threshold. The amplitude of trigger signals
and the range of trigger levels is limited:
±
5 screen divisions
with a channel as trigger source;
±
0.5 V with EXT as trigger
source;
±
5 V with EXT/10 as trigger source; and Inactive with
Line as trigger source. The trigger sensitivity is better than a
third-of-a-screen division.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...