8
–27
Dropout Trigger
Dropout Trigger (
Fig. 8–20
) provides triggering whenever the
signal disappears for a selectable period of time. The trigger is
generated at the end of the time-out period following the “last”
trigger source transition. Time-outs of between 25 ns and 20 s
can be selected.
Figure 8–20. Dropout Trigger Menu (see page 8–37)
Applications
Dropout Trigger is useful for detecting interruptions in data
streams such as network hang-ups and microprocessor crashes.
A typical application is on the last ‘normal’ interval of a signal
that has disappeared completely. This is essentially a single-
shot application, usually with a pre-trigger delay. RIS acquisition
is not useful here because the timing of the trigger timeout is
insufficiently correlated with the input channel signals.
Dropout
Trigger after
timeout if no edge
occurs on
With
slope
Within
Time (25 ns to 20 s)
Positive|Negative
1|2|3|4|Ext|Ext10
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...