C
–13
Using FFT
Error Messages
One of these FFT-related error messages may be displayed at
the top of the screen.
Error Messages
Message
Meaning
“Incompatible input record type”
FFT power average is defined only on a function
defined as FFT.
“Horizontal units don't match”
FFT of a frequency-domain waveform is not available.
“FFT source data zero filled”
If there are invalid data points in the source waveform
(at the beginning or at the end of the record), these are
replaced by zeros before FFT processing.
“FFT source data
over/underflow”
The source waveform data has been clipped in
amplitude, either in the acquisition — gain too high or
inappropriate offset — or in previous processing. The
resulting FFT contains harmonic components which
would not be present in the unclipped waveform. The
settings defining the acquisition or processing should be
changed to eliminate the over/underflow condition.
“Circular computation”
A function definition is circular (i.e. the function is its
own source, indirectly via another function or
expansion). One of the definitions should be changed.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...