10
–4
Zoom, Mathematics and Math Setup
The weight of ‘old’ waveforms in the continuous average
gradually tends to zero (following an exponential rule) at a rate
that decreases as the weight increases.
Summed Averaging
Summed Averaging is the repeated addition, with
equal
weight,
of successive source waveform records. If a stable trigger is
available, the resulting average has a random noise component
lower than that of a single-shot record.
Whenever the maximum number of sweeps is reached, the
averaging process stops. An even larger number can be
accumulated simply by changing the number in the “SETUP”
“for” menu. However, the other parameters must be left
unchanged or a new averaging calculation will be started.
The process may be interrupted by changing the trigger mode
from NORM to STOP or by turning off the active trace using the
respective buttons. Averaging will be resumed when the inverse
is done. The accumulated average is reset by pushing the
CLEAR SWEEPS button or changing an acquisition parameter
such as input gain, offset, coupling, trigger condition, timebase
or bandwidth limit. The number of current averaged waveforms
of the function or its expansion is shown in the Displayed Trace
label.
When summed averaging is selected, the display is updated at a
reduced rate — about once every 1.5 s — in order to increase
the averaging speed (points and events per second).
Summed averaging can also be done on sequence waveforms
to give the average of the segments. And it can be applied to an
expansion of a sequence segment, giving the segment’s
average waveform over many sequence acquisitions.
Extrema Waveforms
These are computed by a repeated comparison of successive
source waveform records with the already-accumulated extrema
waveform, which consists of a maxima (roof) and a minima
(floor) record. Whenever a given data point of the new
waveform is greater than the corresponding maximum roof-
record value, or less than the corresponding floor value, it
replaces it. Thus the maximum and the minimum envelope of all
waveform records is accumulated.
Roof and Floor records can be displayed individually or together.
Содержание 9300C Series
Страница 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Страница 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Страница 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Страница 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Страница 92: ......
Страница 111: ...11 3 Dual Quad ...
Страница 112: ...11 4 Display Parameter s XY only ...
Страница 113: ...11 5 XY Single XY Dual ...
Страница 127: ...12 10 UTILITIES Protected will be displayed on the upper part of the grid whenever the medium is accessed for writing ...
Страница 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Страница 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Страница 142: ...13 2 WAVEFORM STORE RECALL ...
Страница 182: ......
Страница 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Страница 200: ...A 26 Specifications FourChannelsCombined by PP094 Adapter Peak Detect OFF 4 GS s 400k 2M 8M One PP094 input ...
Страница 252: ...E 6 Appendix E ...
Страница 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...