Interval
Maintenance work
monthly
Inspection the drain tap for leaks by visual inspection from the outside
Inspection of the external hoses for material fatigue and leaks
Inspection of the hose clips for correct and secure fit
Inspection of the low level safety function (only for devices with heater)
Cleaning of the condenser (only for air-cooled devices)
Cleaning of the water filter (only for water-cooled devices)
quarterly
Decalcifying of the cooling water circuit (only for water-cooled devices)
A shorter interval must be selected, depending on water hardness and operating time
half-yearly
Inspection of the heat transfer liquid
7.3 Cleaning device
Personnel:
n
Operating personnel
WARNING!
Ingress of cleaning agents in the device
Electric shock
Use a moist cloth for cleaning.
Also note the following:
n
Only clean the control panel with water and detergent. Do not
use acetone or solvents. The consequence would be perma-
nent damage of the plastic surfaces.
7.4 Check low level protection
If the liquid level in the device drops so far that the heater is no
longer completely covered with liquid, an alarm sounds. The dis-
play shows
low level
. The components of the device, the cooling
unit, the heater and the pump are switched off by the electronics.
There must be an alarm signal as soon as low level is
reached.
Maintenance
Variocool
80