Command
Meaning
RMP_IN_01
Query of the current segment number
RMP_IN_02
Query of the set program loops
RMP_IN_03
Query of the current program loop
RMP_IN_04
Query of which program additional commands refer to
RMP_IN_05
Query of which program is currently running (0 = none)
LOG_IN_00_XXXX
Query of a measuring point XXXX from data logger (response: e. g.
020.00_021.23_030.50 => set temperature = 20.00 °C, bath temperature
= 21.23 °C, external temperature = 30.5 °C)
LOG_IN_01
Query of all measuring points from data logger. Unlike with command
"LOG_IN_00", a tab is used as a delimiter here instead of "_". The meas-
uring points are separated with CR and LF. The end is indicated with CR
LF CR LF.
LOG_IN_02
Query of starting time from data logger (response: e. g. 20_14_12_20 =>
day 20, 14:12:20 )
LOG_IN_03
Query of recording interval from data logger (response in seconds)
Please note:
n
For “_” “.” (space character) is also permitted.
n
Unless otherwise specified for the command, the reply is
always in fixed decimal format "XXX.XX" or "-XXX.XX" for neg-
ative values or "ERR_X". (RS 485 interface e.g.
"A015_XXX.XX" or "A015_ERR_X").
n
The command from the control centre must be terminated with
CR, CRLF or LFCR.
n
The response from the constant temperature equipment is
always terminated with CRLF.
n
After every command sent to the constant temperature equip-
ment, it is necessary to wait for a response before the next
command can be sent. This way, the allocation of queries and
responses is clear.
CR = Carriage Return (hex: 0D); LF = Line Feed (hex: 0A)
4.4.6 Write commands of serial interfaces
A write command is a command from the control centre to the con-
stant temperature equipment.
Before starting up
Variocool
34
Содержание VC 1200
Страница 47: ...Fig 13 Menu structure part 1 Operation Variocool 47...
Страница 48: ...Fig 14 Menu structure part 2 Menu structure for Graph Clock and Standby Operation Variocool 48...
Страница 103: ...General Variocool 103...
Страница 110: ......
Страница 111: ......