Personnel:
n
Operating personnel
1.
Select the menu item
Edit
for the selected program.
2.
You can now edit the segments.
Personnel:
n
Operating personnel
Note the following:
n
No time specification is possible in the starting segment. The
temperature of the first segment is reached as quickly as pos-
sible to change to segment 2 after reaching the set tolerance.
n
If the value "0" is entered in the fields
hh
and
mm
, the setpoint
is applied immediately and the bath temperature reached as
quickly as possible.
n
If a tolerance range that is too small has been selected in the
Tolerance
field, the program may not be continued as the
required tolerance is never reached.
n
The default setting for contact modules is
off
. The entry "- -" for
contact modules stands for no changes to the previous seg-
ment, i.e. if "- -" is in all fields, the contact position of the
starting setting or the setting before the program start is main-
tained.
1.
You have the following options:
n
You can display additional columns of the program using
the RIGHT and LEFT arrow buttons.
n
You can navigate in the segments of a program using the
UP and DOWN arrow buttons.
n
You can edit a selected segment by pressing ENTER.
You can adjust the value using the UP and DOWN arrow
buttons. Individual digits can be selected with the RIGHT
and LEFT arrow buttons. Confirm your changes with the
ENTER button.
Fig. 64: Editing program
Editing segments
Operation
Variocool