Code
English output
Description
306
TO1 out of range (Klixon)
Injection temperature outside value range
307
Invalid parameter
Inadmissible parameter in memory
308
CAN system
Problem during internal data exchange
309
Unknown module
Unknown module connected
310
SW Control too old
Software version of control system too old
311
SW Safety too old
Software version of safety system too old
312
SW Command too old
Software version of command remote control unit too old
313
SW Cool too old
Software version of cooling module too old
314
SW Analogue too old
Software version of analogue module too old
315
SW Serial too old
Software version of serial interface (RS232) too old
316
SW Contact old
Software version of contact module too old
317
SW Valve 0 old
Software version of solenoid valve 0 too old
318
SW Valve 1 old
Software version of solenoid valve 1 too old
319
SW Valve 2 old
Software version of solenoid valve 2 too old
320
SW Valve 3 old
Software version of solenoid valve 3 too old
321
SW Valve 4 old
Software version of solenoid valve 4 too old
322
SW Pump 0 old
Software version of pump 0 too old
323
SW Pump 1 old
Software version of pump 1 too old
324
SW Pump 2 old
Software version of pump 2 too old
325
SW Pump 3 old
Software version of pump 3 too old
326
SW HTC old
Software version of high temperature cooler too old
327
SW Ext. Pt100 old
Software version of external Pt100 too old
328
SW Ethernet old
Software version of Ethernet module too old
329
SW EtherCAT old
Software version of EtherCAT too old
341
sm0 min too small
Starting value of injection valve too low
344
chiller missing
Refrigerant unit does not run
345
Valve not closed
Valve of the cooling circuit does not close
347
Configure EEV0
Contact LAUDA Service
348
Configure EEV1
Contact LAUDA Service
349
Preheat unit
A warning concerning possible damage to the cooling
system is sent when the device is operated below 5 °C.
The cooling system must be preheated to prevent
damage. This is automatically performed when outside
installation is activated.
355
CAN buff. overflow
Buffer overflow for CAN reception
Faults
Variocool
89