7.8
Special settings
7.8.1 Setpoint resolution
This function enables the resolution of the setpoint T
set
to be increased from the standard value of
0.1 °C to 0.01 °C (only Master).
Master
S
3x
to
MNEnu
to
PArA
1x
Call setpoint resolution
S
.
The current setting is displayed (here
=1
for the works setting of
0.1 °C).
The resolution display
=1
flashes.
or
=1
for 0.1 °C.
=01
for 0.01 °C.
Wait 4 seconds or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with the Enter key.
7.8.2 Defining the type of start mode
Usually it is desirable that the thermostat carries on operating again after an interruption in the voltage
supply. However, if for safety reasons you do not wish this, you can insert an intervening manual activa-
tion step.