Proline Edition X heating thermostats
YACE0099 / 10/9/2018
43
Command
Control variable
Open the device parameter menu with the
soft key Menu .
With the cursor keys, change further to:
Control
Control variable
.
Intern
is currently active.
Select other control variables (only dis-
played when present) with
or
and
confirm with
or End ,
or quit the window with without making
any changes.
7.7.5 Current consumption from the mains
If your mains fuse is rated below 16 A, the current consumption can be reduced in steps from 16 A to
10 A using this function. Of course the maximum heating power of 3.5 kW is then also reduced accord-
ingly. Take into account whether other loads are still connected to the fused circuit or whether your Pro-
line Thermostat is the only load.
or
Change temperature probe source:
1
for the internal probe,
EP
only when an external probe is connected,
EA
only when an analog module is connected and configured,
ES
only when a serial module is connected and is continuously re-
ceiving actual values from a PC.
Wait 4 seconds
or
Display flashes 4 s
new value is automatically accepted,
or
value is immediately accepted with Enter key.
If
EP
EA
or
ES
has been selected,
then the green
LED indicates that the control has regulated to
the external temperature signal.
Pump Menu
End
T
set
T
fix
Intern
Extern Pt100
Analog module
Digital module
Control variable
12!45
°C
Содержание PROLINE Edition X P 12 C
Страница 2: ......
Страница 4: ......
Страница 15: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Controls and functional elements 15 Proline P 50 C ...
Страница 109: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Technical data 109 Pump characteristics measured with water ...
Страница 112: ...Proline Edition X heating thermostats 112 YACE0099 10 9 2018 Blank page ...
Страница 114: ......
Страница 115: ......