Proline Edition X heating thermostats
YACE0099 / 10/9/2018
63
7.10.3 Interrupting, continuing or terminating the program (Hold, Continue, Stop)
Command
Programmer
Program 1
Status
After a program has been started by
pressing the
key, the command op-
tions
Hold
or
Stop
are shown.
Here, with the aid of the keys
or
and
the running program can be
paused with
Hold
or terminated with
Stop
.
Once the program has been terminated,
the device runs with the last setpoint set-
ting.
Continuation of a program paused with
Hold
occurs using
Continue
which is
obtained with
.
In addition, the standby key
stops the
programmer! Pump and heater are
switched off.
Follow the safety information (
After pressing the standby key
again,
the programmer returns to the previously
selected operating mode:
Pause or active operation depending on
what was previously selected.
Pump Menu
End
Prog. 1 running
Status
Edit
Loops
Graph
Info
Hold
Stop
Pump Menu
End
Prog.1 Standby
Status
Edit
Loops
Graph
Info
Continue
Stop
Pump Menu
End
Prog.1 Standby
Status
Edit
Loops
Graph
Info
Continue
Stop
Содержание PROLINE Edition X P 12 C
Страница 2: ......
Страница 4: ......
Страница 15: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Controls and functional elements 15 Proline P 50 C ...
Страница 109: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Technical data 109 Pump characteristics measured with water ...
Страница 112: ...Proline Edition X heating thermostats 112 YACE0099 10 9 2018 Blank page ...
Страница 114: ......
Страница 115: ......