Proline Edition X heating thermostats
YACE0099 / 10/9/2018
Interface modules
91
8.3.2 Protocol RS-232
The interface operates with 1 stop bit, no parity bit and 8 data bits.
Transfer rate either: 2400, 4800, 9600 (factory setting) or 19200 baud as selected.
The RS-232 interface can be operated with or without hardware handshake,
(RTS/CTS).
The command from the computer must be terminated with CR, CRLF, or LFCR.
The response of the thermostat is always terminated with CRLF.
After each command sent to the thermostat, it is necessary to wait for the reply be-
fore sending another command. This ensures that the sequencing of inquiries and
answers is clear.
CR = Carriage Return (Hex: 0D)
LF = Line Feed (Hex: 0A)
Example:
Transfer of setpoint 30,5 °C to the thermostat
Computer
Thermostat
„OUT_SP_00_30.5“CRLF
„OK“CRLF
8.3.3 Connecting cable RS-485
Thermostat
9-pin sub-D-socket
Pin
Data
1
Data A (-)
5
SG (Signal Ground)
optional
6
Data B (+)
Use screened connecting cables.
Connect screen to connector case.
The connections are galvanically isolated from the rest of the electronics.
Any pins not in use must not be connected!!
An
RS-485
bus always requires bus termination
in the form of a termination network which en-
sures a
defined rest status in the high-resistance
phases of bus operation. The bus termination is
as follows:
This termination network is usually incorporated on the PC plug-in card (RS-485).
Содержание PROLINE Edition X P 12 C
Страница 2: ......
Страница 4: ......
Страница 15: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Controls and functional elements 15 Proline P 50 C ...
Страница 109: ...Proline Edition X heating thermostats YACE0099 10 9 2018 Technical data 109 Pump characteristics measured with water ...
Страница 112: ...Proline Edition X heating thermostats 112 YACE0099 10 9 2018 Blank page ...
Страница 114: ......
Страница 115: ......