Integral T process thermostats
21/01/2019 / YAWE 0026
Preparations
21
6
Preparations
6.1
Assembly and setting up
It is advisable to set up the equipment so that the control unit is towards the front and that the
ventilation for the refrigeration unit, especially on equipment with air-cooled condenser, (ventilation
grille in lower part) is not impeded. Ensure a minimum spacing of 0.5 m between ventilation grille and
wall.
Lock the front castors where appropriate.
Check that the drain cock is closed.
If necessary open the overflow at the back of the equipment and place a vessel underneath it. For
this operation turn the closing plug approx. 45° anticlockwise until the hose end ears can be pulled
out through the openings in the housing.
The overflow is required where it is possible for larger volumes of the heat transfer liquid in the
external circuit to drain back.
45°
Hose end ears