![Lauda Integral T 1200 Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/lauda/integral-t-1200/integral-t-1200_operating-instructions-manual_4012673087.webp)
Integral T process thermostats
21/01/2019 / YAWE 0026
Starting up
87
Message
Explanation
03
Can not write to data memory.
04
Data memory contains data errors.
05
Break or short-circuit of temperature sensor.
06
Temperature measurement circuit does not respond.
07
Measurement circuit for analogue inputs does not respond.
08
Message from control system: temperatures of safety and control systems not identical.
09
Message from safety system: temperatures of safety and control systems not identical.
10
No communication between safety and control systems.
11
Program error safety system.
12
Break of safety temperature sensor.
13
Program error control system.
14
Data error control system.
15
System error control system.