Integral T process thermostats
21/01/2019 / YAWE 0026
Starting up
35
7.5.1
Setpoint selection
(Level 0)
or
Press keys until SET (Setpoint) appears.
Enter, display is flashing.
or
Enter setpoint with both keys (
Section 7.3.1).
Display is flashing 4 s
new value is entered
automatically,
or
value entered immediately with key.
For safety reasons the setpoint can only be set up to 2 °C above the upper limit of
the operating temperature range of the particular equipment type.
Appears when analogue setpoint input is activated.
7.5.2
External actual temperature display
switches line 2 of the display from setpoint SET to the
external actual temperature T
E
,
or internal actual temperature (outflow temperature) T
I
if
external control is activated.