Integral T process thermostats
21/01/2019 / YAWE 0026
Starting up
61
Analogue setpoint input is switched on, with current input
(I) configuration. With voltage the display shows AIS U.
Scroll to AIE
or
pressing the key leads to submenu setpoint, “Analogue
inputs” (
Section 7.6.5.1).
AIE = analogue Input external actual temperature as
standard signal/ instead of Pt100 on input 10S.
OFF = this input is switched off, the measurement of the
Pt100 at socket 10S is valid.
Input is switched on. With current configuration (I). In case
of voltage the display shows AIE U.
Pressing the key leads to the submenu “analogue inputs”
external actual temperature (
Section 7.6.5.1)
or
A01 = analogue output 1
OFF = switched off.
Analogue output 1 is switched on. With voltage output (U)
configured and the internal temperature signal, i.e. outflow
temperature (t I).
I = current output
tE = external actual temperature
tS = setpoint
P = pressure
Y = electrical output
Outputs 2 and 3 are similar to output 1.