Integral T process thermostats
22
Preparations
YAWE 0026 / 21/01/2019
Water-cooled version
The heat of condensation and the motor heat are dissipated through a water-cooled counter flow heat
exchanger. Connect up the hoses. The connection for inlet (from water tap) and outlet (to drain) are
located on the back of the equipment at the bottom. Inlet on the left, outlet to drain on the right, looking
on the back of the equipment. The cooling water flow is automatically adjusted to the required amount
through the condenser pressure.
Secure the hoses with hose clips to prevent slipping off!!
The water consumption depends on the energy to be dissipated. It ranges between 200 and 2000 L/h
depending on equipment type, cooling water temperature and loading. The cooling water temperature
must not exceed 25 °C. Cooling water pressure should be
>
2.5 bar. The maximum overpressure must
not exceed be 10 bar.
Connection of the cooling water
Note that the following conditions apply for the connection of the cooling water supply:
Cooling water pressure (feed - outlet)
maximum 10 bar overpressure
Differential pressure
p (feed - outlet)
minimum 2.5 bar
Cooling water temperature
10 to 15 °C recommended,
10 to 25 °C not exceed (with power restrictions)
Cooling water quantity
see Technical Data (
Cooling water hose for connection to the
device
minimum 13 mm
Water inlet
Water outlet