11/8/00
Using ID Setting Files
7-13
R+A -
The Logic Output turns on when an RMS
exceedance event occurs or when an event occurs
on one of the A:D channels.
P+A -
The Logic Output turns on when a Peak
exceedance event occurs or when an event occurs
on one of the A:D channels.
R+P+A -
The Logic Output turns on when either
an RMS or Peak exceedance event occurs or when
an event occurs on one of the A:D channels.
I+A -
The Logic Output turns on when each inter-
val begins or when an event occurs on one of the
A:D channels.
R+I+A -
The Logic Output turns on when an
RMS exceedance event occurs, when each interval
begins, or when an event occurs on one of the A:D
channels.
P+I+A -
The Logic Output turns on when a Peak
exceedance event occurs, when each interval
begins, or when an event occurs on one of the A:D
channels.
R+P+I+A -
The Logic Output turns on when
either an RMS or Peak exceedance event occurs,
when each interval begins, or when an event
occurs on one of the A:D channels.
R/S -
The Logic Output turns on when the instru-
ment is in run mode and turns off when the instru-
ment is in stop mode.
Excd - The Logic Output turns on when a valid
exceedance is detected and after it’s minimum
duration has been met.
Logic Output Time Setting -
(SSA, LOG, ISM)
The Logic Output Time sets the number of seconds that the
logic output will remain turned “ON” following a triggering
event. Accordingly, this setting can take on any integral
value within the following range:
(0)sec to (255)sec
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...