10-16
824 Reference Manual
11/8/00
The RT60-A and RT60-B RTA instruments are accessed
from the
S
SETUP key.
RT60-A
Although the settings are predefined
in RT60-A and RT60-B, they can be
changed at any time and used as cus-
tomized setups for a personalized
RT60 measurement.
The RT60-A setup is designed to gather data to measure an
RT60 by exciting a room with a steady state noise. It is con-
figured in the following way:
•
Detector: Linear
•
Detector Time: 0.01 seconds
•
Autostore: Yes
•
PreTrigger: 50
•
Save ByTime: Yes
•
Triggering: ON
•
Trigger Source: 1 kHz
•
Arm Mode: L >
•
Arm Level: 80.0 dB
•
Arm Time: 2.0 second
•
Trigger Mode: L <
•
Trigger Level: 80.0 dB
•
Trigger Time: 00.0000 seconds
•
End Mode: Time
•
End Time: 3.0 seconds
•
Then: Stop
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...