8-12
824 Reference Manual
9/13/00
you to create a custom report for any SSA or LOG instru-
ment ID.
The Tailored Report includes the Data Report, Log Report,
Setup Report, Interval Report (Short), Exceedance Report,
and the Daily Report.
After selecting the template you can modify it as explained
in “Printing a Custom Report” on page 8-22.
Tailored Report Template (SSA instrument)
Report Settings
Lower Level
Report Settings
Description
Unformated [No]
A dump of all Read, Query and Interval remote commands. Multiple
values are comma delimited.
Data Report [Yes]
This shows general information such as Leq, SEL, Dose, Projected Dose,
Lmin, Lmax, Peak, Ln’s, exceedance, and calibration logs.
Setup Report [Yes]
Print the settings and their values for the SSA instrument.
Intv Report [Short]
Leq, SEL, Lmin, Lmax, Peak, Uwpk, exceedances, time, date, and dura-
tion in the short report. The Long reports adds the Ln values and Spectral
LNs for the Interval period.
Hist Report [No]
Leq for each time history period.
Print Range
Intv Report [Short]
You can customize the interval report by choosing the long or short
report.
Lowest Record (00000)
Selects the lowest interval record to be printed.
Highest Record (65535)
Selects the highest interval record to be printed
Hist Report [No]
You can customize the time history report by choosing the long or short
report.
Lowest Record (00000)
Selects the lowest interval record to be printed.
Highest Record (>99999)
Selects the highest interval record to be printed
Histograms
SPL Histogram [No]
Prints SPL histogram and overall Spectral LN report.
Lowest Level (045)
Selects the lowest SPL histogram level to be printed.
Highest Level (095)
Selects the highest SPL histogram level to be printed.
Resolution (5.0)
Selects the SPL histogram resolution
Pk-1 Histogram [No]
You can customize the Peak1 Histogram report by choosing the long or
short report.
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...