2-16
824 Reference Manual
9/12/00
Step 8
To verify that the instrument is properly cali-
brated, use the
u
key to highlight “Check”.
Step 9
Place the calibrator over the microphone, switch
the calibrator ON, and press
c
to activate the cal-
ibration Check process.
The display will show the measured level and the difference
from the calibrator level entered and the measured level as
shown.
Step 10
If the calibration is not reasonably close to the
level of the calibrator, the calibration can be
changed by choosing Change. Press
c
, use
the
d
key to highlight “Change”.
Step 11
Switch the calibrator ON, and press
c
.
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...