5-4
824 Reference Manual
9/12/00
With this instrument the SLM and RTA functions are operat-
ing simultaneously. The primary indicator range of the SLM
function is 80 dBA, sufficient for most applications.
In operation, the instrument measures four sets of frequency
spectra; the running instantaneous spectrum, the L
eq
spec-
trum, the spectrum associated with the maximum SPL level
that occurred during the measurement and the minimum bin
by bin spectrum. It also gives you a time history record and
Ln centiles. The time history record stores up to 38 parame-
ters of your choice, including RTA Leq, RTA Max, or RTA
Min, which is determined by the Time History setup under
the settings menu.
SSA Measuring Ranges
By using a 24-bit DSP processor, the digital filters of the
System 824 meets the specifications of IEC 61260 - Class 1
and ANSI S1.11-1986 Type 1-D over a linearity range
exceeding 80 dB. The dynamic range of the analysis func-
tion, from noise floor to overload, is 85 dB (typical). Eight
ranges are available, providing gain to +50 dB and attenua-
tion to -20 dB in 10 dB steps.
The following presents the SPL, Peak Detectors and 1/3
Octave Filter Measuring Ranges for the SSA instrument:
A-weighting, C-weighting, and Flat are measured to
ANSI S1.4 and IEC 60651 standards for SLM linearity.
1/3 octave filters listed are measured to the IEC 61260
standard for Class 0 linearity which is ±0.3 dB. The num-
bers in the table represent SPL levels with a typical 2541
microphone which has 44.1 mV/Pa sensitivity.
Measurement Range (SSA mode only)
0dB Gain
10dB Gain
20db Gain
30dB Gain
40dB Gain
50dB Gain
NF
Start
End
Range
NF
Start
End
Range
NF
Start
End
Range
NF
Start
End
Range
NF
Start
End
Range
NF
Start
End
Range
AW
33
48
128
80
24
38
118
80
15
28
108
80
8
23
98
75
7
20
88
68
6
20
78
58
CW
32
48
128
80
23
38
118
80
16
28
108
80
9
23
98
75
8
20
88
68
8
20
78
58
FLT
32
48
128
80
25
38
118
80
21
33
108
75
21
33
98
65
26
33
88
50
23
40
78
38
16.0
2
45
128
83
2
35
118
83
1
25
108
83
2
25
98
73
-1
28
88
60
1
23
78
55
31.5
3
43
128
85
1
35
118
83
1
23
108
85
2
25
98
73
-4
22
88
66
-4
23
78
55
63.0
4
40
128
88
0
33
118
85
1
23
108
85
0
22
98
76
-7
15
88
73
-7
18
78
60
125
2
40
128
88
-2
27
118
91
-3
28
108
80
-2
18
98
80
-9
11
88
77
-8
8
78
70
250
3
37
128
91
-3
25
118
93
-3
15
108
93
-4
15
98
83
-10
8
88
80
-9
8
78
70
500
12
33
128
95
4
25
118
93
-3
15
108
93
-4
13
98
85
-10
6
88
82
-10
8
78
70
1000
10
32
128
96
2
20
118
98
-4
15
108
93
-5
13
98
85
-10
6
88
82
-10
8
78
70
2000
15
32
128
96
6
20
118
98
-2
15
108
93
-5
13
98
85
-8
8
88
80
-8
8
78
70
4000
14
32
128
96
7
20
118
98
0
15
108
93
-2
13
98
85
-6
8
88
78
-6
8
78
70
8000
14
32
128
96
7
20
118
98
1
15
108
93
-1
13
98
85
-2
12
88
76
-3
13
78
65
*NF is noise floor
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...