12-14
824 Reference Manual
11/8/00
Pulse/FM-b Screen
To access the Pulse/FM-b screen, press the VIEW key
,
then select Pulse/FM with the up/down arrows and press the
check key
.
This will display the Pulse/FM-a screen. Press the left or
right arrow
to view the Pulse/FM-b screen.
This screen displays required values for the calibration of
the frequency modulated tone capabilities of an audiometer.
Present the frequency modulated tone at an appropriate level
and press the RUN/STOP key
. The stability of the mea-
surement can be evaluated visually. To retrieve data, press
RUN/STOP
again and read the values from the screen.
The following values are displayed: carrier frequency, maxi-
mum frequency, minimum frequency, modulation rate, all in
hertz.
“ xxxx ” indicates that the carrier has
been stable for 1 second.
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...