9/12/00
A -3
For example, the System 824’s Logging SLM Overall-a dis-
play shows the current SPL (Sound Pressure Level) in 1/
10th decibel (dB) resolution. The actual real-time resolution
is 1/32nddB. The displayed SPL is followed by the text
“dB” and then the frequency weighting in effect—set to A in
this example.
The System 824’s Logging SLM Overall-e display shows
the overall measurement period and TWA. In this case the
measurement is Leq (because of the 3dB exchange rate).
The quantity is A-weighted and has a measurement elapsed
time of 11:43.1 (11 hours, 43 minutes, and 1 second).
The actual equations used within the analyzers are based on
that for IEC-804 and are implemented according to this
equation:
Elapsed measurement time, shown as mm:ss.ss,
hh:mm:ss or hhhhh:mm
Scrolling SPL history
Current SPL while running or at the time a stop
occurred
The currently-selected frequency weighting for
SPL
The currently-selected detector
Elapsed measurement time, shown as mm:ss.ss,
hh:mm:ss or hhhhh:mm
Overall time weighted average (L
eq
in this case)
Time-weighted average (TWA) descriptor for
the overall measurement
Overall sound exposure level (SEL)
Overall exchange rate
Содержание System 824
Страница 4: ...824 Reference Manual iv...
Страница 18: ...824 Reference Manual xiv...
Страница 34: ...1 16 824 Reference Manual 11 8 00...
Страница 88: ...3 38 824 Reference Manual 11 8 00...
Страница 118: ...5 10 824 Reference Manual 9 12 00...
Страница 180: ...6 62 824 Reference Manual 11 8 00...
Страница 314: ...9 36 824 Reference Manual 11 8 00...
Страница 356: ...10 42 824 Reference Manual 11 8 00...
Страница 371: ...11 2 00 FFT Optional 11 15...
Страница 372: ...11 16 824 Reference Manual 11 2 00...
Страница 388: ...12 16 824 Reference Manual 11 8 00...
Страница 393: ...9 12 00 A 5 The following data indicates which settings are associated with which TWA data variables...
Страница 398: ...A 10 824 Reference Manual 9 12 00...
Страница 485: ...9 13 00 C 17 Windscreen Corrections...
Страница 503: ...9 13 00 C 35 Graph 20 824 in the YZ plane fitted with the 2541 micro phone 12 5 kHz...
Страница 504: ...C 36 824 Reference Manual 9 13 00...
Страница 536: ...824 Reference Manual x...