Glossary UH50 – FOR INTERNAL USE ONLY!
Gyr GmbH
Glossary UH606-101b
Page
34
of
38
Terminals (module)
To connect the external cables to the module, 2-way or 4-way (Phoenix) screw terminals are
used. Here is an extract from the data sheet:
Insulation stripping length
−
5
mm
Connection capacity
−
rigid or flexible, 0.2 - 2.5 mm²
−
flexible with wire-end ferrule, 0.25 - 1.5 mm²
−
conductor sizes 26 - 14 AWG
Multiple-conductor connection
(2 conductors with same cross-section)
−
rigid or flexible, 0.2 - 0.75 mm²
−
flexible with wire-end ferrule without plastic sleeve, 0.25 - 0.34 mm²
−
flexible with TWIN ferrule with plastic sleeve, 0.5 - 0.75 mm²
Recommended screwdriver
−
0.6 x 3.5 mm
Tightening torque
−
0.4
Nm
Test mode
Test mode can also be called from normal mode by telegram (“heat meter to Pb”).
Test mode
From here it is possible to call any of the test modes.
Return to normal mode is performed automatically after no more than 15 hours.
Test mode flowrate
The flowrate measurements are performed in a measurement timebase of 250ms. Pulses
proportional to the volume are output via the optical interface.
The flowrate is – depending on volume measuring unit - displayed with a resolution of e.g.
0.0001 m³/h (100 ml/h). Each update of the display value is indicated by brief display of an
arrow symbol on the LCD.
Measured flowrate
Return to normal mode is performed automatically after no more than 15 hours.
In the
test mode menu
, it is possible to call the required test mode manually. The display loop
can be operated with LCD button 1. The required function is called by pressing LCD button 2.
The values for volume and quantity of heat measured in normal mode remain unchanged.
While the calibration seal is not set, the required test mode can also be called by telegram via
the optical interface.
Call
test mode volume
Call
test mode flowrate
Call
test mode quantity of heat
Call
test mode temperature
Return to normal mode
Test mode is ended by pressing LCD button 1. A star symbol (*) now appears in the second
digit as acknowledgment. When LCD button 1 is pressed again, the next line of the menu is
shown and it is possible to advance further through the display loop.
Return to normal mode is performed automatically after no more than 15 hours.
Test mode quantity of heat
The accumulated quantity of heat is displayed with a resolution of 0.01 kWh. The average of
640 temperature measurements is evaluated.
The measurements are ended when the simulated volume of 2 m³ is reached. Then the numeric
values for the measured quantity of heat and the simulated volume are displayed at appox. 2-
second intervals.