Glossary UH50 – FOR INTERNAL USE ONLY!
Gyr GmbH
Glossary UH606-101b
Page
16
of
38
A special version of the pulse module (“Deluxe”) is available with an Opto-MOS output.
Application example: Battery-powered pulse receiver.
Advantages: Low voltage drop and polarized (bipolar).
Labeling pulse
module
Type
open collector
Voltage
maximal 30 V =
Current
maximal 30 mA
Dielectric strength 500 V
rms
against ground
Classification
OB (per EN 1434-2)
Voltage drop
approx. 1.3 V at 20 mA
Classification
OC (per EN 1434-2)
Voltage drop
approx. 0.3 V at 0.1 mA
With the
M-bus module
, several heat meters
can be read from a central station.
The data frame can be defined with a jumper.
Jumper in pos. 1 “Fixed data frame”
Jumper in pos. 2 “Guaranteed data frame”
Jumper in “center” “Variable data frame”
M-bus module
Po
s.
2
Po
s.
1
Jumper
M-Bus
Note: If two M-bus modules are plugged, please note the
restrictions
!
Display on LCD
MB
(meter bus)
Standard
per EN 1434-3
Isolation galvanic
Connection
with galvanic isolation
Voltage
50 V maximum
Current consum. 1.0 M-bus loads (1.5 mA)
Addressing
primary or secondary
Transmission rate 300/1200/
2400
/4800 baud
Literature TKB
3417
Read-out freq.
any number of time, also with battery
Data scope with fixed data frame:
- Update after every 15 minutes
Property number; quantity of heat; volume; setting and resetting a user lock; setting the data
and time, and setting the primary and secondary address.
Data scope for variable/guaranteed data frame:
- Update after every 15 minutes *)
Property number; unit number; manufacturer’s identifier; medium; firmware version, fault
messages; missing time; operating time; quantity of heat; volume; previous month’s values for
quantity of heat, tariff register, volume, maximum for power and missing time; previous year’s
values for quantity of heat, volume and maximum for power ; measuring period with maxima for
power, flowrate and temperatures; actual values for power, flowrate and temperatures; F0
prewarning; setting and resetting a user lock; setting the date and time, and setting the primary
and secondary address.
*) Special version with minimum update rate of 30 s can be supplied (WZU-MB-30), but it has a
shorter battery life
Data scope for fast read-out:
- Update after a minimum of 4 s
Property number; quantity of heat; volume; actual values for power, flowrate and temperatures,
and setting the date and time.
Fast read-out via the M-Bus is only permissible with the appropriate battery (D cell for 6 years)
or a power supply module.
The
CL module
can be used to set up a point-to-point
link enabling the heat meter to be read remotely, for
example, at the front door.
Display on LCD
CL
(current loop)
Standard
per EN 1434-3
Type
passive current loop
CL module
RTX+,
RTX-