1 Installation
10
QuickGene-810 Operation Manual
Rev. 1.7
■
Setup the Pump Tube
The following explains how to setup the Pump Tube.
(1) Turn the power on.
Press the [1] side of the switch.
When the QuickGene-810 has been initialized, the
LCD displays "DNA WHOLE BLOOD" (this name
can be changed). It takes about a minute.
(2) Press the [DISCHARGE] button.
Press the [DISCHARGE] button. This operation
makes the pump rotate to the position for ease to
setup pump tubes.
(3) Turn off the power.
Press the [O] side of the switch.
(4) Open the front cover.
Hold the handle and open the front cover until the
upper limit is reached.
(5) Attach the Pump Tube to the pump.
Attach the Pump Tube to both the Wash Buffer
Pump and the Elution Buffer Pump.
Note: When setting the Pump Tube, wear gloves.
CAUTION:
The front cover will not stay
on the way to the upper
limit. If you release the front
cover while opening it, it
may cause injury or device
damage.
Important:
When attaching the Pump Tube, wear
gloves. Do not touch the Drive Head in
the pump without gloves.
Power switch
ON
Wash Buffer
Pump
Elution
Buffer Pump
Drive Head
After removal of
the pump cover
Содержание QuickGene-810
Страница 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Страница 2: ......
Страница 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Страница 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Страница 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Страница 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Страница 76: ......