1
2
6
3
4
5
7
5 How to Set Parameters
38
QuickGene-810 Operation Manual
Rev. 1.7
The following table shows the meaning of each parameter.
Display
sequence
LCD message
Parameter name
Meaning
1
BIND PEAK: 50 -
BIND PEAK: 160
Peak pressure on the
sample
Sets the peak pressure during pressure application. Judgment
criteria is "the pressure before pressure appli the
pressure set in this item". Pressure application may be
stopped before reaching the peak pressure, if air pressure
drops suddenly, indicating that all liquid has been discharged.
Can be specified between 50 - 160 kpa in increments of 1kpa.
2
WASH COUNT: 0 -
WASH COUNT: 5
Frequency of wash
buffer discharges
Sets the frequency of wash buffer discharges.
Can be specified between 0 - 5 times.
3
WASH PEAK: 50 -
WASH PEAK: 160
Wash buffer peak
pressure
Sets the peak pressure during pressure application. Judgment
criteria is "the pressure before pressure appli the
pressure set in this item". Pressure application may be
stopped before reaching the peak pressure, if air pressure
drops suddenly, indicating that all liquid has been discharged.
Can be specified between 50 - 160 kpa in increments of 1kpa.
4
WASH VOL1: 250 -
WASH VOL1: 750
1st Wash Buffer
Discharge Rate
Sets the first wash buffer discharge rate.
Can be specified between 250 - 750µl in increments of 50l.
5
WASH VOL2: 250 -
WASH VOL2: 750
2nd WBDR
Sets the second wash buffer discharge rate.
Can be specified between 250 - 750µl in increments of 50l.
6
WASH VOL3: 250 -
WASH VOL3: 750
3rd WBDR
Sets the third wash buffer discharge rate.
Can be specified between 250 - 750µl in increments of 50l.
7
WASH VOL4: 250 -
WASH VOL4: 750
4th WBDR
Sets the fourth wash buffer discharge rate.
Can be specified between 250 - 750µl in increments of 50l.
8
WASH VOL5: 250 -
WASH VOL5: 750
5th WBDR
Sets the fifth wash buffer discharge rate.
Can be specified between 250 - 750µl in increments of 50l.
9
WASH DIP TM: 0 -
WASH DIP TM: 510
Wash buffer
immersion
management timer
Sets the wash buffer immersion management time.
Can be specified between 0 - 510 sec. in increments of 2
seconds.
10
WAS2 WAIT T: 0 -
WAS2 WAIT T: 255
Waiting time
Sets the time to wait for shifting to wash buffer discharges in
the second round.
Can be specified between 0 - 255 min. in increments of 1
minute.
11
WAS2 COUNT: 0 -
WAS2 COUNT: 5
Frequency of wash
buffer discharges in
the 2nd round
Sets the frequency of wash buffer discharges in the second
round.
Can be specified between 0 - 5 times.
12
WAS2 PEAK: 50 -
WAS2 PEAK: 160
Wash buffer peak
pressure in the 2nd
round
Sets the peak pressure during pressure application. Judgment
criteria is the pressure before pressure appli the
pressure set in this item. Pressure application may be stopped
before reaching the peak pressure, if air pressure drops
suddenly, indicating that all liquid has been discharged.
Can be specified between 50 - 160 kpa in increments of 1kpa.
13
WAS2 VOL1: 250 -
WAS2 VOL1: 750
1st WBDR in the 2nd
round
Sets the first wash buffer discharge rate in the second round.
Can be specified between 250 - 750µl in increments of 50l.
14
WAS2 VOL2: 250 -
WAS2 VOL2: 750
2nd WBDR in the 2nd
round
Sets the second wash buffer discharge rate in the second
round.
Can be specified between 250 - 750µl in increments of 50l.
15
WAS2 VOL3: 250 -
WAS2 VOL3: 750
3rd WBDR in the 2nd
round
Sets the third wash buffer discharge rate in the second round.
Can be specified between 250 - 750µl in increments of 50l.
16
WAS2 VOL4: 250 -
WAS2 VOL4: 750
4th WBDR in the 2nd
round
Sets the fourth wash buffer discharge rate in the second round.
Can be specified between 250 - 750µl in increments of 50l.
17
WAS2 VOL5: 250 -
WAS2 VOL5: 750
5th WBDR in the 2nd
round
Sets the fifth wash buffer discharge rate in the second round.
Can be specified between 250 - 750µl in increments of 50l.
Содержание QuickGene-810
Страница 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Страница 2: ......
Страница 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Страница 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Страница 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Страница 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Страница 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Страница 76: ......