page 96
Appendix
IV
Mounting steps for a multi-dosing unit if ordered
afterwards
The following steps are usually not necessary if the multi-Dosing unit was
ordered together with the DSA100.
IVa)
Assembling the syringe driver unit(s)
Depending on the number of syringes one or two syringe driver units with two to
four syringe drivers each must be assembled to left and right-hand chamber of the
DSA100.
Overview
The following table shows the required combinations for all multi-dosing units.
dosing unit
syringe driver units
DS3222 (2 syringes)
2-fold driver left
DS3223 (3 syringes)
3-fold driver left
DS3224 (4 syringes)
4-fold driver left
DS3225 (5 syringes)
3-fold driver left
2-fold driver right
DS3226 (6 syringes)
4-fold driver left
2-fold driver right
DS3227 (7 syringes)
4-fold driver left
3-fold driver right
DS3228 (8 syringes)
4-fold driver left
4-fold driver right