4. Select the syringes one after another (except the one marked with
M
) and carry
out the menu order
Reset Position
(DSA1) respectively
Reset Syringe
(DSA3/DSA4) for each syringe. The syringe drivers are moved into their start
positions and initialized.
DSA1
DSA3/4
SPAC SPAC SPAC SPAC
10.2
Inserting the syringes
SPAC SPAC SPAC SPAC
1. Select the first syringe in the DSA1/DSA3/DSA4 software (
Chapter 10.1).
2. In order to remove air from the refill tube the dosing valve must be switched to
channel the liquid from the syringe to the refill tube.
For
DSA1
: Select the tab sheet
Refill
from the
Device Control
panel. Click on
the
Valve Position for Dosing
button. The button label changes to Valve
Position for Refill.
For
DSA3/DSA4
: Open the drop-down menu for syringe selection. Select the
menu item
Tools
Toggle Switch Syringe Valve
. The next time the menu is
opened a dot indicates that the valve is switched to the refill tube.