page 84
Humidity chamber HC10
Thermostat connection
The TC30/40 and the HC10 are connected in series. The thermostat liquid flows
from the male quick-release connector to the female quick-release connector.
Male
Female
1. Connect the output of the through-flow thermostat to the input of the
environmental chamber.
2. Connect the output of the environmental chamber to the input of the HC10.
3. Connect the output of the HC10 to the input of the through-flow thermostat.
Gas connector for humidity control
!
Dry gas must be used for all connections when the humidity of the air used for
measurement is low. If the flow of gas is produced by a compressor then the air
must be dried if the humidity of the air used for the measurement is to be lower
than the atmospheric humidity.
Connect the connector on the throttle valve to a compressed gas connection
(1 bar).
!
Please maintain the specified input pressure of 1 bar.
!
The throttle valve reduces the pressure of the gas to 350 mbar. This is
permanently set and locked at this value. Do not try to adjust the valve.