page 86
Humidity chamber HC10
12.7
Fitting the temperature sensor
Feed the temperature sensor into the opening in the environmental chamber.
Connect the temperature sensor to the connector on the DSA100 provided for
the purpose (
Chapter 5.8).
12.8
Covering the needle inlet
The needle inlet of the environmental chamber is covered in order to guarantee a
good regulation of the humidity. Silicone covers are supplied for this purpose:
three covers with small opening for sessile drop measurements
three covers with large opening for pendant drop measurements
Procedure
1. Feed the silicone cover over the dosing needle as far as it will go.
2. Set the required dosing height.
3. Slide the silicone cover down so that the cover rests on the dosing needle inlet.
SPAC SPAC SPAC SPAC