EN
96200-xxx_00 GA USV Tropic 2.0_20160229.docx
- 63 -
General electronic devices
Malfunction
e.g. stop or change of the intended operation
mode
14.3
Electromagnetic immunity
Guidelines and manufacturer's declarations – electromagnetic immunity
The ultrasonic nebulizer is intended for use in the electromagnetic environment specified below.
The customer or user of the product should ensure that it is operated in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment –
guideline
Electrostatic discharge/
ESD
according to IEC 61000-4-2
±
6 KV
contact discharge
±
8 KV
air discharge
±
6 KV
contact discharge
±
8 KV
air discharge
Floors should be made from wood or concrete or tiled
with ceramic tiles.
If the floor is covered with synthetic material, the relative
humidity level must be at least 30 %.
Electrical fast
transients/bursts
according to IEC 61000-4-5
±
2 KV
for power supply lines
±
2 KV
for power supply lines
Mains voltage quality should be that of a typical
commercial or hospital environment.
Surges on AC mains lines
according to IEC 61000-4-11
±
1 KV
differential mode
voltages
±
2 KV
common mode voltages
±
1 KV
differential mode
voltages
±
2 KV
common mode voltages
Mains voltage quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
fluctuations of the supply
voltage
according to IEC 61000-4-11
>95 % dip, 0.5 periods
60 % dip, 5 periods
30 % dip, 25 periods
>95 % dip, 5 seconds
-100 %
-60 %
-30 %
-100 %
Mains voltage quality should be that of a typical
commercial or hospital environment.
If the user of the ultrasonic nebulizer requires continued
operation even during energy supply interruptions, it is
recommended that the ultrasonic nebulizer is powered
from an uninterruptible power supply or a battery.
Power frequency magnetic
field
(50/60 Hz)
according to IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical commercial or hospital
environment.
Note: U
T
is the alternating mains voltage prior to application of test levels.
14.4
Electromagnetic immunity, devices not intended for life support
Guidelines and manufacturer's declarations – electromagnetic immunity, devices not intended for life support
The ultrasonic nebulizer is intended for use in the electromagnetic environment specified below.
The customer or user of the product should ensure that it is operated in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment –
guideline
Conducted RF disturbances
according to IEC 61000-4-6
Radiated RF disturbances
according to IEC 61000-4-3
3 V
eff
150 kHz – 80 MHz
3 V/m
80 MHz – 2.5 GHz
3 V
eff
150 kHz – 80 MHz
3 V/m
80 MHz – 2.5 GHz
Portable and mobile radio equipment should not
be used at a smaller distance to the product,
including lines, than the recommended separation
distance, which is calculated by means of the
equation appropriate for the transmission
frequency.
Recommended separation distance:
d = 1.2
√P (for 150 kHz
– 80 MHz)
d = 1.2
√P (for 80MHz
– 800 MHz)
d = 2.4
√P (for 800 MHz
– 2.5 GHz)
with P as nominal output of the transmitter in
watts (W) in accordance with the transmitter