Warning, risk of explosion
Do not use the device together with combustible or explosive liquids.
Warning, risk of patient injury
Do not use the ultrasonic nebulizer on patients requiring ventilation
Warning, risk of patient injury
Do not install the ultrasonic nebulizer in breathing systems
Warning, risk of patient injury
Do not use the ultrasonic nebulizer with a ventilation mask, tracheal cannula, or mouthpiece.
This can lead to accumulations of liquid in the respiratory system.
Warning, risk of infection
Do not use the ultrasonic nebulizer without a bacterial filter
The air flow is only filtered adequately if a suitable bacterial filter is used.
Replace the bacterial filter after 24 hours of operation or in accordance with the special instructions
provided by the respective filter manufacturer.
Reusable hoses must be cleaned once a day or whenever there is a change of patient.
Disposable hoses must be replaced once a day or whenever there is a change of patient.
The ultrasonic nebulizer must only be used for its intended purpose. Due to the respirable particle size of the mist,
particular care regarding hygiene must be taken during preparation and execution of the inhalation. Medication should
only be nebulized with a medication container (11.2). A medical professional must determine the need for nebulizing,
especially regarding the inhalation of medication.
5.4
Self-evident improper use
Warning, observe environment of use!
The device must not:
•
be covered, thrown, forcefully opened or dropped,
•
be used for patient ventilation,
•
be used without bacterial filter or fluids,
•
be operated by children who are unsupervised,
•
be used together with combustible or explosive liquids or operated in potentially explosive areas.
No objects must be placed in the housing
5.5
Device components
1 Ultrasonic nebulizer
2 Nebulizer chamber, complete
3 Hose approx. 300 mm
7 Bacterial filter connection
17 Nebulizer chamber
18 Nebulizer chamber lid
1
2
3
4
5
7
1
1
1
2
6