INSTALLATION
3
21
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
3.3.7 Step 7: mount the sensor rails
The best suitable sensor is automatically selected from a list of three available sensor types and
the appropiate number of traverses is indicated.
Choose the right traverse mode. For recommendations please see examples above.
1.7
Transducer set
Calibration number
Number of traverses
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
Ta
522505050
13-04-2010 14:11:09 13 MB free
Transducer set
Fill in the value given on the label on the transducer cable
Calibration number
Fill in the value given on the label on the transducer cable
Number of traverses
Z mode = 1 traverse
V mode = 2 traverses
W mode = 4 traverses
Pipe
Available sensor
Traverse mode
DN15...150
Small 2 MHz, 1 rail
DN50...250
Medium 1 MHz, 1 rail
DN200...750
Medium 1 MHz, 2 rails
DN400...1500
Medium 1 MHz, 2 rails