INSTALLATION
3
13
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
3.2 Installation requirements
3.2.1 Inlet, outlet and recommended mounting area
To perform an accurate flow measurement preferably mount the sensor rail at least 10 DN
downstream of a flow disturbance like elbow, valve, header or pump. Please follow the
installation recommendations in the next installation position examples.
CAUTION!
•
In order to comply with the EMC directive 2004/108/EC, I/O cables that provide a galvanic
connection to the UFC 400 P should have a maximum total length of 3 meter.
•
When not used, put the connector covers of the connectors on the bottom side of the
converter in place. This to prevent improper functioning caused by dust/dirt.
•
When the sensor cables are connected while the converter is positioned on a flat surface,
turn the handle fully backwards (towards the housing) in order to prevent excess stress on
the sensor cables.
•
With an empty main battery, the backup battery of the real time clock may run empty when
this is the case for a period longer than one year.
•
The protection degree of the battery charger / mains adapter is IP 40 / NEMA 1. It should be
protected against moisture entering.
•
To prevent damage due to vibrations, do not firmly attach the converter to or place it on top of
a vibrating object.
Figure 3-1: Inlet, outlet and recommended mounting area
1
Min. 10 DN
2
Min. 5 DN
3
OK, 120
°