TECHNICAL DATA
7
71
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Ultrasonic transit time
Application range
Flow measurement of liquids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Transit time
Secondary measured value
Volume flow, mass flow, flow speed, flow direction, speed of sound, gain,
signal to noise ratio, diagnosis value, reliability of flow measurement,
quality of acoustic signal, thermal energy (requires input of temperature
(2x)).
Design
The measurement system consists of one or two measuring sensors and
a portable signal converter.
Signal converter
Signal converter
Signal converter
Signal converter
Portable housing
UFC 400 P
Measuring sensor
Measuring sensor
Measuring sensor
Measuring sensor
Standard
Single or dual rail sensors with 1 or 2 MHz transducers
Optional
OPTISONIC 6300 flow sensors using cable adapters
Diameter ranges
Diameter ranges
Diameter ranges
Diameter ranges
DN15...150 /
½
...6"
One rail, 2 MHz transducers
Outer diameter must be at least 20 mm /
¾
".
DN50...250 / 2...10"
One rail, 1 MHz transducers
DN200...1500 / 8...60"
Two rails, 1 MHz transducers
Options
Options
Options
Options
Outputs
0(4)...20 mA, pulse, frequency and/or status output
Inputs
0(4)...20 mA (2x) with optional I/O box.
Counters
4 internal counters with a maximum of 8 counter places, for counting
volume, energy and/or mass units.
USB
1x host port (PC can use OPTISONIC 6400 as a removable media device)
1x slave (memory stick can be written by converter)
Self diagnostics
Integrated verification, diagnostic functions: flowmeter, process,
measured value, empty pipe detection, bar graph.