INSTALLATION
3
27
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
3.3.8 Step 8: connect the sensor cable
CAUTION!
Please look at the labels on the cable of each transducer for the calibration number. All
transducers for the same measuring path MUST have the same calibration number!
INFORMATION!
The signal cables are prefixed to the transducers in the factory.
Figure 3-12: Connecting signal cables
1
Connector for "UP" transducer (blue)
2
Connector for "DOWN" transducer (green)
1
2
3
4
5
6
7
8
9
+/-
0
EXP
ABC
DEF
GHI
JKL
MNO
PQRS
TUV
WXYZ
A/a
-
I/O
DISPLAY
LOGGER
MENU
1
1
2
2
1
2
INFORMATION!
After mounting the transducers in the proper position, it may take 45 seconds before a signal is
found.
INFORMATION!
Advice on signal strength:
Advice on signal strength:
Advice on signal strength:
Advice on signal strength:
Signal > 75%:
Signal > 75%:
Signal > 75%:
Signal > 75%: good signal
Signal 50...75%:
Signal 50...75%:
Signal 50...75%:
Signal 50...75%: fairly good signal
Signal 10...50%:
Signal 10...50%:
Signal 10...50%:
Signal 10...50%: low signal
Signal < 10%:
Signal < 10%:
Signal < 10%:
Signal < 10%: bad or no signal, check settings in menu, change transducer distance until there is
at least a low signal