3
INSTALLATION
20
OPTISONIC 6400
www.krohne.com
08/2011 - 4000972602 - MA OPTISONIC 6400 R02 en
3.3.6 Step 6: program the converter
In the menu "Settings and information
Device
Password" a password can be set. In
measuring mode the password blocks the measurement mode after the display sleep time has
passed. The measurement mode can only be left after typing the password.
Select "Installation" from the main menu, then press "
" and select the desired configuration.
CAUTION!
Because the converter remember most of its previous settings, you may wish to reset the
converter before installing at a new location to prevent unexpected behavior. To reset the
converter, enter the "Settings and information" menu and select "Load factory settings".
1.2
Pipe tag
Outer diameter
Material
Wall thickness
Liner material
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
Pipe1
100.00 mm
Carbon steel
5.00 mm
None
13-04-2010 14:11:09 13 MB free
Pipe tag
Give the pipe a name
Press again "
" and "
".
Outer diameter
Use the outside diameter
Material
Choose the right material
Wall thickness
Fill in the pipe wall thickness material
Liner material
Select whether there is a liner or not
Choose next
CAUTION!
Inaccurate input of the outside diameter will affect the accuracy of the measured flow rate.
1.3
Fluid
VoS Fluid
Viscosity
< Previous | Next >
< Previous | Next >
< Previous | Next >
< Previous | Next >
Water
1485.0 m/s
1 mm
2
/s
13-04-2010 14:11:09 13 MB free
Fluid
Give the right liquid from the table. If the liquid is unknown,
choose water. The correct value will be measured.
VoS Fluid
Velocity of Sound. Only change it if it is well known.
Viscosity
Only change it if it is well known.
Choose next
Содержание OPTISONIC 6400
Страница 81: ...NOTES 8 81 OPTISONIC 6400 www krohne com 08 2011 4000972602 MA OPTISONIC 6400 R02 en...
Страница 82: ...8 NOTES 82 OPTISONIC 6400 www krohne com 08 2011 4000972602 MA OPTISONIC 6400 R02 en...
Страница 83: ...NOTES 8 83 OPTISONIC 6400 www krohne com 08 2011 4000972602 MA OPTISONIC 6400 R02 en...