Troubleshooting
Information about faults
Thanks to the high quality standard faults in the electronic unit are very rare. Implausible display values
therefore usually indicate faults in the plant. The following fault table lists the various fault messages as
well as their cause and remedy.
Alarms
Fault display
Cause and elimination
Checked
There is no fault.
►
No operating steps necessary.
Incorrect password input.
►
Repeat the password entry with
the correct password.
The frequencies of Volumeter A
are not entered in ascending order.
►
Enter the frequencies in
ascending order, see "Menu 4:
K-factors Volumeter A", Page 44.
The frequencies of Volumeter B
are not entered in ascending order.
►
Enter the frequencies in
ascending order, see "Menu 5:
K-factors Volumeter B", Page 44.
The temperatures are not entered
in ascending order.
►
Enter the temperatures in
ascending order, see "Menu 6:
Density table 1/Density
determination", Page 44.
The temperatures are not entered
in ascending order.
►
Enter the temperatures in
ascending order, see "Menu 7:
8.00 No alarm.
The device is
working properly.
8.01 Alarm
Password not
valid.
8.02 Alarm
K-factors Vol. A
Frequencies not
in ascending order!
8.03 Alarm
K-factors Vol. B
Frequencies not
in ascending order!
8.04 Alarm
Density table 1
Temperatures not
in ascending order!
8.05 Alarm
Density table 2
Temperatures not
in ascending order!