Connecting cables
C
o
m
m
is
s
io
n
in
g
1. Remove the rear device cover.
2. Adapt the cable entry to the cable diameter and cut it to length, see "Connecting cables to the
tension spring terminals", page 24.
3. Pull the cables for the analog, relay and pulse outputs individually through the cable entries.
4. Connect the cables for the analog, relay and pulse outputs according to the wiring diagram, see
"Connection assignments", page 21.
5. Lay the cables for the analog, relay or pulse outputs to the consumer.
6. Connect the consumer.
7. Replace the rear device cover.
Connecting the power supply
Prerequisites:
□
All sensors are correctly connected
Aids:
□
KRAL tool set
□
Diagonal cutter
1. Remove the rear device cover.
2. Adapt the cable entry to the cable diameter and cut it to length, see "Connecting cables to the
tension spring terminals", page 24.
3. Ensure that the system is de-energized.
4. Pull the supply cable (24 V DC) through the cable entry.
5. Connect the supply cable (24 V DC).
6. Connect the supply cable (24 V DC) to the power supply of the system.
7. Replace the rear device cover.
The electronic unit is ready to operate.
CAUTION
Damage to equipment through incorrect wiring of analog or pulse outputs.
►
Do not supply the analog or pulse outputs with voltage (active outputs!).
CAUTION
Damage to equipment through incorrect wiring of consumer inputs.
►
Before connecting the electronic unit to the power supply, ensure that all consumers (temperature
sensors, pick ups) are connected correctly, see wiring diagram.