Customer Support
1.800.366.5343
–
20. Set your techniques to
70kV / 7mAs
and take your
fourth
exposure.
21. If the exposure was acceptable, click
Next
to move to the next frame.
22. Continue to take exposures with
techniques set to
70kV / 7mAs
for the
remaining frames (5-19).
23. After taking all 19 defect exposures,
click the
Back
button if you want to
retake any previous exposures, then
click
Prep
and retake the exposure.
Otherwise, if you are satisfied with all
exposures, click the
Next
button in the
top right corner.
24. Wait for the
Defect Map Generation
to
complete and then click
Next
.
25. Calibration is complete.
Close
the
window by clicking
X
in top right corner.